The model H2 – WLD 1000 is a simple, accurate and inexpensive hydrogen measurement in the trace and percentage range.
A clearly arranged display and calibration allows an easy and fast installation of the model.
Parameters | Hydrogen % |
Method | Thermal conductivity |
ATEX | Nein |
SIL | Nein |
Range | 0 – 5 % |
Accuracy | +/- 3 % FSD |
Response time T 90 | 2s |
Output | 4 – 20 mA |
Supply Voltage | 10 – 35V |
One user-specific measuring range
Measuring ranges between 0.5% and 100% are available.
Sensor
In the H2 – WLD 1000 a thermal conductivity cell is used to measure hydrogen in the sample gas, depending on the measuring range.
The sensors set industrial standards for accuracy, sensitivity, ease of use and lifetime.
Calibration
The calibration of the measuring device is done with a calibration gas.
This gas can be freely selected within the measuring range.
Advantages
Stainless steel cell housing
The cell housing has a modular design and is made of stainless steel
Parameters | Hydrogen % |
Method | Thermal conductivity |
Range | 0 - 5 % |
ATEX | Nein |
SIL | Nein |
Accuracy | +/- 3 % FSD |
Resolution | 0,1 % |
Flow rate | ca. 1 l/m |
Pressure | 0,1 - 20 bar |
Calibration | Calibration Gas |
Response time T 90 | 2s |
Gas temperature | 0 - 50 °C |
Output | 4 - 20 mA |
Supply Voltage | 10 - 35V |
Display | Graphical Display |
Protection Class | IP 65 |
Connection | 6 mm |
Dimensions | 145 x 80 x 85 |
Weight | 2 Kg |
Your satisfaction is our top priority. That is why we have made it our mission to offer our customers a comprehensive portfolio. We are happy to conduct detailed consulting and coordination discussions in order to find optimal solutions for your concerns together in cooperation with you. Through our years of experience in gas analysis and moisture measurement we are your ideal partner.