Home > Oxymaster II 16P

Oxymaster II 16P

Measuring Percent Oxygen in Gas

The Model Oxymaster was made for fast, accurate and economic measurement of Oxygen at trace levels.
A simple menu-driven software, touchpad and large LCD-display allow easy and fast start-up of the instrument.

Parameters Oxygen %
Method

Electro chemical

ATEX Nein
SIL Nein
Range

0 – 100 %
0 – 25 %
0 – 5 %
freely programmable

Accuracy

+/- 2 % FSD

Response time T 90

0 – 10 % = 30s

Output

4 – 20 mA
RS232
Alarm
Datalogger

Supply Voltage

85 – 230 V

User-selectable measuring range

The user can select any range within the 0 – 10000 ppm.
Free-programmable concentration alarm and a system alarm give versatility meeting almost any requirement.

 

Data outputs and Data logger

For the Oxygen concentration is a 4-20 mA output standard available. Other outputs are available on request. RS232 for data reading is standard. A data logger with 10000 values is included.

 

Sensor with long operating life

The Oxymaster uses a special fuel cell to measure the oxygen concentration. The sensor meets the industrial requirements for accuracy,
sensitivity, easy to use and operating life.

 

Calibration

The calibration of the instrument for trace oxygen measurements in gas, should be done with a calibration gas. The concentration can be
chosen freely within the measuring range

 

Customer

The Oxymaster can be used as a stand alone unit, but a complete, custom made, system with sample handling is also supplied by us.
Optional it is possible to have the Oxymaster as two channel unit with a second measuring cell.

 

Benefits

  • Compact economic instrument with sensor separate from the electronics
  • Sensor with long operatiing life.
  • No range swichting requiredInstructions and error messages in normal text (eg calibration not OK, change mesuring cell)

 

Flow-through measuring cell in Stainless Steel

  • The measuring cell is modular and is made of stainlesss steel
  • A defective measuring cell can be repaired by replacing the defective part only, rather than the complet unit
Parameters Oxygen %
Method

Electro chemical

Range

0 – 100 %
0 – 25 %
0 – 5 %
freely programmable

ATEX Nein
SIL Nein
Accuracy

+/- 2 % FSD

Resolution

0,01 %

Flow rate ca. 1 l/m
Pressure 0,1 – 1 bar
Calibration

Air
Calibration Gas

Response time T 90

0 – 10 % = 30s

Gas temperature 0 – 50 °C
Output

4 – 20 mA
RS232
Alarm
Datalogger

Supply Voltage

85 – 230 V

Display

Graphic display

Protection Class IP 65
Connection 6 mm
Dimensions 220 x 190 x 190
Weight 3 Kg
  • Variety of sensors available or can be used
  • User programmable measuring range
  • Output signal: 4-20 mA for O2
  • Output signal: 4-20 mA for Temp. (opt.)
  • one oxygen alarm 2 A
  • Selfdiagnosis with relay output
  • Clear text graphic display
  • Data tracking, temperature reading
  • SS316 measuring cell
  • Instructions and error messages in normal text
  • universal Powersupply 85 – 230 VAC 50 – 60 Hz
  • 24 VDC / ca. 50 Watt
  • two channel unit optional sensor block optional with 11/4” tread
  • Semi conductor industry
  • Gas manfacturers
  • Metalurgical industries
  • Chemical inductry

Kontaktieren Sie uns

Contact Us