The Model RH CAL is a microprocessor based, programmable humidity calibration system that is at home in the metrology lab or out in the field performing on site NIST traceable humidity calibrations. The system offers the highest accuracy available for both relative humidity and ambient temperature.
Parameters | Moisture % Temperature °C Dewpoint °C |
Method | Dew Point Mirror |
ATEX | Nein |
SIL | Nein |
Range | At 25°C |
Accuracy | +/- 0,5 % rH |
Response time T 90 | 1°C / sec |
Output | 4 – 20 mA |
Supply Voltage | 100 V – 220 V |
Unlike other comparative systems RH CAL is entirely selfsufficient. It does not need compressed air or a water connection to operate, which allows this system to be truly portable.
With the RH CAL, temperature and relative humidity are controlled independently; therefore you are not limited to performing calibrations at the surrounding ambient temperature, which may not be appropriate for your unique
calibration protocol. Using EdgeTech’s Optical Chilled Mirror (OCM) primary measurement technique for traceability and feedback control, RH CAL is a standalone, portable humidity calibration system.
Continuing to RH CAL’s ease of use are features such as maintenance reducing Automatic Balance Cycle (ABC), and integral ambient temperature probe and D2 chilled mirror sensor which are located in the sample chamber providing not only superior accuracy but the fastest response.
The Model RH CAL employs a unique control scheme for maintaining such precise RH control. The system incorporates volumetric proportional control valves. By independently modulating the “Dry” and “Wet” valves from full-open to full-close,
and all points between, any desired humidity can be quickly generated and stabilized upon. Additionally, because the RH CAL is continuously monitoring the sample chamber conditions via its integral chilled mirror, it is able to instantaneously react to changes in humidity and maintain control stability.
Parameters | Moisture % Temperature °C Dewpoint °C |
Method | Dew Point Mirror |
Range | At 25°C |
ATEX | Nein |
SIL | Nein |
Accuracy | +/- 0,5 % rH |
Resolution | 0,1 °C |
Flow rate | 0,25 – 2,5 l/min |
Pressure | entfällt |
Calibration | Moisture Generator |
Response time T 90 | 1°C / sec |
Gas temperature | 10°C – 50°C |
Output | 4 – 20 mA |
Supply Voltage | 100 V – 220 V |
Display | Graphic Display |
Protection Class | IP 65 |
Connection | 6mm |
Dimensions | 52 x 20 x 43 cm |
Weight | 5 Kg |
Your satisfaction is our top priority. That is why we have made it our mission to offer our customers a comprehensive portfolio. We are happy to conduct detailed consulting and coordination discussions in order to find optimal solutions for your concerns together in cooperation with you. Through our years of experience in gas analysis and moisture measurement we are your ideal partner.